accQcomm Interface Module
Section 2 Installation and Operation
2-12
Note
For the accQmin flow meter, the accQcomm displays and
transmits secondary level data at 10X the actual value when
using Standard Data format. This should be considered when
setting up alarm conditions and analog outputs. This will not
occur with an accQmin using Reduced Data format.
Testing Analog Outputs To verify that the analog outputs are functioning properly.
1. Attach the leads of a current meter to the output circuit to
be tested.
2. Click Test Analog Outputs to open the test dialog.
Figure 2-10 Analog output test dialog
3. Select a COM port and baud rate for communication.
4. Select the channel and enter a current value in mA.
5. Click Test Output.
The measured current value should match the set value,
until or unless changed by another event or action.
Page view 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 63 64

Comments to this Manuals

No comments

Blue Sea Systems PN7520 manuels

Manuels d'utilisation et guides de l'utilisateur pour Chargeurs de batterie Blue Sea Systems PN7520.
Nous fournissons des manuels en pdf 1 Blue Sea Systems PN7520 à télécharger gratuitement par type de document : Manuel de l'utilisateur






Plus de produits et de manuels pour Chargeurs de batterie Blue Sea Systems

Modèles Type de document
7521 (25A) Manuel de l'utilisateur   Blue Sea Systems 7521 (25A) User manual, 32 pages
7521 (25A) Manuel de l'utilisateur   Power Conversion, 6 pages
7522(40A) Manuel de l'utilisateur   Blue Sea Systems 7522(40A) User manual, 32 pages